Basic principles:
Spectrophotometer analysis is based on the selective absorption of electromagnetic radiation by substances with different molecular structures.
The application of electromagnetic radiation spectrum region is usually ultraviolet-visible spectrophotometer in 200nm~800nm.
The use of microprocessor control technology to make qualitative and quantitative analysis of substances in the UV-visible spectrum region is a necessary multi-purpose analytical instrument in conventional laboratories.
Features:
The holographic blazed grating Monochromator has the advantages of high wavelength accuracy, good monochromism and low stray light.
Large screen with backlit liquid crystal display, measurement data at a glance.
Using microcomputer measurement system, Tmura conversion accuracy is high, and there are automatic adjustment of 0% T and 100% T, concentration factor setting, concentration direct reading.
The measurement has high accuracy, good reproducibility and stability.
Optional serial printer and use with software UVWin7 to expand the function of the instrument.
Main technical indicators:
Wavelength range; 200nm~1000nm.
Large allowable error of wavelength: ±2nm.
Wavelength repeatability: ≤ 1nm.
Large allowable error of transmittance: ±0.5% (T) (measured by NBS930D).
Renaturation of specific gravity of transmission: ≤ 0.2% (T).
Spectral bandwidth: 4nm.
Stray light: ≤ 0.3% (T) (at 220nm, measured by NaI, at 360nm, measured by NaNO2).
Stability: dark current drift: 0.2 (T) / 3min bright current drift: 0.5 (T) / 3min.
Other:
Instrument size and weight: 500mm × 350mm × 190mm 14kg
752N 752NP1us N2 N2S N4 N4S L6 L6S L7 L8 L9 760CRT L5 L5S 754N L3 L3S 754N N4 N4S 752G